Equipment report

sim-freezer-009 · generated 2026-08-14T22:47

Health Score
94/100
Status
EXCELLENT
Confidence
86%

Sentinel analysis

Critical degradation relative to the equipment baseline has been detected.

Recommendation: Inspect sensor wiring, connectors, probe placement and acquisition electronics before diagnosing the refrigeration system.

Technical findings

FindingSeverityConfidenceExplanation
INTERNAL TEMP STABLEINFO95%Internal temperature trend is stable.
INTERNAL TEMP VARIABILITYMEDIUM80%Internal temperature variability has reached 0.79 °C.
COMPRESSOR TEMP STABLEINFO95%Compressor temperature trend is stable.
COMPRESSOR CURRENT STABLEINFO95%Compressor current trend is stable.
INTERNAL TEMP BASELINE OKINFO95%Internal temperature remains consistent with this equipment's baseline. Current average: -78.653 °C; baseline: -78.661 °C.
COMPRESSOR TEMP BASELINE OKINFO95%Compressor temperature remains consistent with this equipment's baseline. Current average: 51.602 °C; baseline: 51.646 °C.
COMPRESSOR CURRENT BASELINE OKINFO95%Compressor current remains consistent with this equipment's baseline. Current average: 1.203 A; baseline: 1.203 A.
EVAPORATOR DELTA BASELINE OKINFO95%The internal-to-evaporator temperature difference remains consistent with this equipment's baseline.
EVAPORATOR DELTA STABLEINFO92%Cooling temperature difference remains stable.
COOLING EFFORT BASELINE OKINFO92%Compressor thermal and electrical effort remain consistent with the equipment baseline.
COMPRESSOR TEMPERATURE C IMPOSSIBLE JUMPHIGH90%Compressor temperature sensor changed by 23.206 °C between two consecutive measurements. The provisional maximum accepted jump is 15.000 °C.