sim-freezer-005 · generated 2026-08-14T22:46
Critical degradation relative to the equipment baseline has been detected.
Recommendation: Inspect sensor wiring, connectors, probe placement and acquisition electronics before diagnosing the refrigeration system.
| Finding | Severity | Confidence | Explanation |
|---|---|---|---|
| INTERNAL TEMP STABLE | INFO | 95% | Internal temperature trend is stable. |
| INTERNAL TEMP STABILITY OK | INFO | 95% | Internal temperature variability remains within the provisional nominal range. |
| COMPRESSOR TEMP STABLE | INFO | 95% | Compressor temperature trend is stable. |
| COMPRESSOR CURRENT STABLE | INFO | 95% | Compressor current trend is stable. |
| INTERNAL TEMP BASELINE OK | INFO | 95% | Internal temperature remains consistent with this equipment's baseline. Current average: -80.009 °C; baseline: -80.005 °C. |
| COMPRESSOR TEMP BASELINE OK | INFO | 95% | Compressor temperature remains consistent with this equipment's baseline. Current average: 57.505 °C; baseline: 52.862 °C. |
| COMPRESSOR CURRENT BASELINE OK | INFO | 95% | Compressor current remains consistent with this equipment's baseline. Current average: 1.204 A; baseline: 1.208 A. |
| EVAPORATOR DELTA BASELINE OK | INFO | 95% | The internal-to-evaporator temperature difference remains consistent with this equipment's baseline. |
| EVAPORATOR DELTA STABLE | INFO | 92% | Cooling temperature difference remains stable. |
| COOLING EFFORT DRIFT | MEDIUM | 82% | A moderate increase in compressor effort has been detected relative to this equipment's baseline. |
| COMPRESSOR TEMPERATURE C IMPOSSIBLE JUMP | HIGH | 90% | Compressor temperature sensor changed by 27.918 °C between two consecutive measurements. The provisional maximum accepted jump is 15.000 °C. |